Development of an Automatic Seed Marker Registration Algorithm Using CT and kV X-ray Images
Kwang Ho Cheong, Byung Chul Cho, Sei Kwon Kang, Kyoung Joo Kim, Hoon Sik Bae, Tae Suk Suh
1Department of Radiation Oncology, Hallym University College of Medicine, Korea.
2Department of Biomedical Engineering, College of Medicine, The Catholic University of Korea, Seoul, Korea. suhsanta@catholic.ac.kr
ABSTRACT
PURPOSE:
The purpose of this study is to develop a practical method for determining accurate marker positions for prostate cancer radiotherapy using CT images and kV x-ray images obtained from the use of the on-board imager (OBI).
MATERIALS AND METHODS:
Three gold seed markers were implanted into the reference position inside a prostate gland by a urologist. Multiple digital image processing techniques were used to determine seed marker position and the center-of-mass (COM) technique was employed to determine a representative reference seed marker position. A setup discrepancy can be estimated by comparing a computed COMOBI with the reference COMCT. A proposed algorithm was applied to a seed phantom and to four prostate cancer patients with seed implants treated in our clinic.
RESULTS:
In the phantom study, the calculated COMCT and COMOBI agreed with COMactual within a millimeter. The algorithm also could localize each seed marker correctly and calculated COMCT and COMOBI for all CT and kV x-ray image sets, respectively. Discrepancies of setup errors between 2D-2D matching results using the OBI application and results using the proposed algorithm were less than one millimeter for each axis. The setup error of each patient was in the range of 0.1+/-2.7~1.8+/-6.6 mm in the AP direction, 0.8+/-1.6~2.0+/-2.7 mm in the SI direction and -0.9+/-1.5~2.8+/-3.0 mm in the lateral direction, even though the setup error was quite patient dependent.
CONCLUSION:
As it took less than 10 seconds to evaluate a setup discrepancy, it can be helpful to reduce the setup correction time while minimizing subjective factors that may be user dependent. However, the on-line correction process should be integrated into the treatment machine control system for a more reliable procedure.
Key Words: kV x-ray image, On-board imager, Seed marker, Prostate, Registration
TOOLS
PDF Links  PDF Links
Full text via DOI  Full text via DOI
Download Citation  Download Citation
Share:      
METRICS
1,607
View
29
Download
Related article
Editorial Office
Department of Radiation Oncology, Samsung Medical Center,
Proton Therapy Center, B2, 81, Irwon-ro, Gangnam-gu, Seoul 06351, Republic of Korea
Tel : +82-2-3410-3617
E-mail: rojeditor@gmail.com, roj@kosro.or.kr
Copyright © The Korean Society for Radiation Oncology.                      Developed in M2PI
Close layer
prev next