| Home | E-Submission | Sitemap | Contact Us |  
top_img
J Korean Soc Ther Radiol > Volume 6(2); 1988 > Article
Journal of the Korean Society for Therapeutic Radiology 1988;6(2): 263-268.
A Study on Dose Distribution around Fletcher-Suit Colpostat Containing 137Cs Source
Wee Saing Kang
Department of Therapeutic Radiology, College of Medicine, Seoul National University, Seoul, Korea.
ABSTRACT
This paper presents dose distributions in water around Fletcher-Suit colpostat containing 137Cs tube, and shielding effect of internal lead shield. Using ready packed film, author measured dose distribution in water around the colpostat containing cesium source. Nine sheets of films on one side of the colpostat are packed with acryl frames cut out so as to fill water, and irradiated in water by cesium source in the colpostat. Dose distributions on transverse plane and upper plane 0.5 cm from upper surface of the colpostat were measured. Shielding effect was greater in upper medial direction than in lower medial direction. And that was the greatest around 30degree from the axis of the colpostat on upper side and around 50degree on lower side. In the region 7 cm from the center of the colpostat, shielding efficiency was 0.23 to 0.35 on the lower 50degree and 0.26 to 0.42 on the upper 30degree, and decreased with increase of distance.
Key Words: Dose distribution, Fletcher-Suit colpostat, 137Cs source
TOOLS
PDF Links  PDF Links
Full text via DOI  Full text via DOI
Download Citation  Download Citation
Share:      
METRICS
1,589
View
24
Download
Related article
A Study on Dose Distribution around Fletcher-Suit Colpostat Containing Cs-137 Source by a Computer1989 December;7(2)
Editorial Office
Department of Radiation Oncology, Samsung Medical Center,
Proton Therapy Center, B2, 81, Irwon-ro, Gangnam-gu, Seoul 06351, Republic of Korea
Tel : +82-2-3410-3617
E-mail: rojeditor@gmail.com, roj@kosro.or.kr
Copyright © The Korean Society for Radiation Oncology.                      Developed in M2PI
Close layer
prev next