Dosimetric comparison of intensity-modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) in total scalp irradiation: a single institutional experience
Article information
Abstract
Purpose
Total scalp irradiation (TSI) is a rare but challenging indication. We previously reported that non-coplanar intensity-modulated radiotherapy (IMRT) was superior to coplanar IMRT in organ-at-risk (OAR) protection and target dose distribution. This consecutive treatment planning study compared IMRT with volumetric-modulated arc therapy (VMAT).
Materials and Methods
A retrospective treatment plan databank search was performed and 5 patient cases were randomly selected. Cranial imaging was restored from the initial planning computed tomography (CT) and target volumes and OAR were redelineated. For each patients, three treatment plans were calculated (coplanar/non-coplanar IMRT, VMAT; prescribed dose 50 Gy, single dose 2 Gy). Conformity, homogeneity and dose volume histograms were used for plan.
Results
VMAT featured the lowest monitor units and the sharpest dose gradient (1.6 Gy/mm). Planning target volume (PTV) coverage and homogeneity was better in VMAT (coverage, 0.95; homogeneity index [HI], 0.118) compared to IMRT (coverage, 0.94; HI, 0.119) but coplanar IMRT produced the most conformal plans (conformity index [CI], 0.43). Minimum PTV dose range was 66.8% –88.4% in coplanar, 77.5%–88.2% in non-coplanar IMRT and 82.8%–90.3% in VMAT. Mean dose to the brain, brain stem, optic system (maximum dose) and lenses were 18.6, 13.2, 9.1, and 5.2 Gy for VMAT, 21.9, 13.4, 14.5, and 6.3 Gy for non-coplanar and 22.8, 16.5, 11.5, and 5.9 Gy for coplanar IMRT. Maximum optic chiasm dose was 7.7, 8.4, and 11.1 Gy (non-coplanar IMRT, VMAT, and coplanar IMRT).
Conclusion
Target coverage, homogeneity and OAR protection, was slightly superior in VMAT plans which also produced the sharpest dose gradient towards healthy tissue.
Introduction
Extensive superficial malignancies of the scalp constitute a rare but challenging clinical presentation in oncology. Most common malignancies treated with total scalp irradiation (TSI) include angiosarcoma, cutaneous lymphoma and squamous cell carcinoma [1]. However, radiotherapy of the scalp can be technically demanding due to the unique clinical presentation of the tumor on the one side (multifocal spread, irregular borders) and the complex geometry of the scalp on the other side (convex surface, close proximity of visual organs and face).
In addition to the anatomical challenges, intense treatment planning efforts, complex and staff-intensive treatment and patient setup (immobilisation) further complicate this clinical situation. Radiotherapeutic approaches to treat the entire scalp range from electron-based over photon-based to combined techniques including intensity-modulated radiotherapy (IMRT), volumetric-modulated arc therapy (VMAT), tomotherapy and high dose rate (HDR) brachytherapy [2-5].
However, sufficient data to determine the optimal and most effective radiation regimen is still lacking and frequently, both the individual and specific clinical tumor presentation as well as the technical conditions at the institution limit the scope of possible techniques to be chosen.
Achieving an adequate dose coverage in the target volume, that is the dermis and epidermis, while sparing brain parenchyma was historically attempted by matched, shifted or overlapping electron fields [6-8]. However, these techniques not only suffered from significant dose inhomogeneity but also from electron scatter complicating dosimetry [9]. These obstacles could be partially resolved by using photon beam IMRT techniques or photon-electron combination approaches which resulted in improved target dose coverage, dose homogeneity and minimized difficulties with doses at field junctions [4,9-11].
We previously reported our institution’s experience and comparatively evaluated coplanar and non-coplanar photon IMRT in the TSI of scalp malignancies under a dosimetric perspective and found that non-coplanar IMRT techniques were superior in dose homogeneity, target volume coverage and sparing of the optical structures [12].
Novel motion-based radiation techniques such as non-static photon IMRT and VMAT in particular, promise superior dose distribution and conformity [13,14]. Recently, first evidence showed encouraging clinical response in a patient with extensive leukaemia cutis of the head and scalp treated by VMAT [15].
However, a robust analysis comparing static-field IMRT with VMAT has not been undertaken. This retrospective treatment planning study now evaluates the dosimetric merits of IMRT against VMAT and comparatively assesses plan conformity index (CI), homogeneity index (HI), dose-volume relation and organ-at-risk (OAR) protection in radiotherapy of the whole scalp.
Materials and Methods
1. Radiation treatment plan calculation and characteristics
A patient databank search was performed at the Department of Radiation Oncology of the Martin Luther University Halle-Wittenberg and 5 patients who previously received whole brain radiotherapy (WBRT) for cerebral metastasis (without prior cranial surgery) were randomly selected. All clinical and imaging information was analysed anonymously. This specific patient group (i.e., with intracerebral metastasis but without malignant scalp lesions) was chosen to ensure homogeneous and identical planning conditions with an anatomically intact and unaltered scalp surface to best evaluate the dosimetric differences of the radiation techniques. Generally, the previous WBRT was based on computed tomography (CT)-images obtained in treatment position (GE LightSpeed RT; without contrast) with the patients’ head immobilized in an individual thermoplastic mask without use of a bolus. It was carried out in a three-dimensional (3D) conformal radiation technique.
Treatment plans were restored and the entire prior plan content including previously delineated target volumes and OAR contours were deleted. For each patient, planning target volume (PTV) (i.e., from skin surface to skull cap), OAR (brain, optic chiasm, left and right optic nerve, left and right lens, brainstem) were re-delineated by an experienced radiation oncologist using the Oncentra MasterPlan external beam planning software (Nucletron, Columbia, MD, USA) (Fig. 1). For each patient, three treatment plans were generated using the collapsed cone algorithm: a coplanar IMRT plan, a non-coplanar IMRT plan and a VMAT plan, based on the same PTV of each individual patient, yielding a total of 15 radiation treatment plans for comparison. For IMRT plans, 20 iteration steps per cycle were applied for creating the fluency and 80 iteration steps for optimization and consecutive segmentation of the multi-leaf-collimator. Planning time for IMRT plans ranged from 2 to 4 hours per plan which were calculated as an 11-field step-and-shoot technique. Non-coplanar IMRT used 6 couch angles ≠ 0° (45°, 315°, 90) and VMAT plans were based on 4 arcs each at 180° (180°–340° clockwise, 20°–180° clockwise, 180°–20° counter clockwise, 340°–180° counter clockwise), with arcs covering the same angle range being a dual arc (Fig. 2). A 5-mm virtual bolus (without air gap between bolus and scalp) was created and used for all treatment plans to ensure sufficient surface dose. Plan calculation time for VMAT ranged from 6 to 8 hours per plan.
Dose prescription for all plans was 50 Gy to the PTV (whole scalp) at fractions of 2 Gy (6 MV). Radiation doses are reported as maximum dose (Dmax), D2%, minimum dose (Dmin) and D98% and dose gradient across the target volume is reported in Gy per mm. Doses at OARs are reported as mean dose in Gy (brain and lenses) or maximum dose in Gy (optical nerves, brain stem, optic chiasm), respectively. For the brain dose constraint, a maximum average dose was used and dose constraints for the PTV were Dmin (D95%) 47 Gy, Dmin (D98%) 48.5 Gy, Dmean 50 Gy, and Dmax 54 Gy. OAR doses were lowered as much as reasonable achievable, that is until a significant deterioration of the plan homogeneity and PTV coverage occurred. Monitor unites (MUs) of each treatment was also recorded to evaluate the execution efficiency.
2. Dose conformity
In the literature, different conformity indexes were published [16]. In this study, the conformity index (CI) as defined by Paddick [17] was used:
where TV(PIV) refers to the target volume covered by the prescription isodose volume, and TV is the target volume. PIV describes the prescription isodose volume, referring to the 95% or 98% isodose. The homogeneity index (HI) was defined according to the ICRU report 83 [18-20] as
with D50 being the median absorbed dose and D2 and D98 representing the doses received by 2% and 98% of the PTV. PTV coverage was the ratio of the PTV receiving 95% of the prescribed dose (V95%) and the corresponding PTV:
Results
1. Dosimetric analysis IMRT and VMAT treatment plans
Dosimetric comparison of the coplanar and non-coplanar IMRT and VMAT plans including PTV coverage, homogeneity, conformity, minimum and maximum doses are presented in Table 1 for each individual patient case. Mean MUs for all VMAT plans were 950.8 followed by non-coplanar IMRT plans (962.8) and coplanar IMRT (974.2). Treatment plan delivery was 10–12 minutes for IMRT and 2–3 minutes for VMAT plans on average.
Mean dose gradient was 1.6, 1.1, and 1.3 Gy/mm in VMAT, coplanar and non-coplanar IMRT. PTV coverage was 0.94 for coplanar/non-coplanar IMRT and 0.95 for VMAT plans (mean). Mean HI was 0.119 for IMRT and 0.118 for VMAT plans while the mean CI was 0.45 (CI98) in non-coplanar, 0.4 in coplanar IMRT and 0.43 in VMAT plans (CI95 was 0.39 for coplanar IMRT and 0.43 for non-coplanar IMRT and VMAT). Minimum PTV dose (mean) was 43.7 Gy (VMAT), 41.4 Gy (non-coplanar IMRT) and 41.3 Gy (coplanar IMRT); maximum PTV dose (mean) was 54.7 Gy (VMAT, non-coplanar IMRT) and 54.8 Gy (coplanar IMRT). D98 was 46.5, 46.4, and 46.6 Gy and D2 was 52.2, 52.3, and 52.5 Gy in coplanar, non-coplanar IMRT and VMAT.
Except for 1 patient (#2), VMAT produced the sharpest dose gradients of all plans. In 2 patients, PTV coverage was significantly higher in VMAT plans (0.95, patient #2 and 0.97, patient #3); while in 2 patients, TV coverage was comparable between treatment plans (patients #4 and #5) and in 1 patient (#1), PTV coverage was superior in the non-coplanar IMRT. HI was comparable between treatment plans (Table 1).
In all patients, non-coplanar IMRT plans were more conformal than coplanar IMRT which applies to 3 out of 5 patients for VMAT.
All treatment plans produced dose hot spots within and outside the target volume, however, they were comparable between treatment plans of each individual patient as were maximum doses which ranged from 107.2–112.6 in coplanar IMRT, from 106.6–112 in non-coplanar IMRT and from 104.8–113.3 in VMAT plans which also produced both the lowest (104.8, patient #5) and the highest maximum dose (113.3, patient #1). Minimum PTV doses ranged from 66.8%–88.4% in coplanar IMRT plans, from 77.5%–88.2% in non-coplanar IMRT plans and from 82.8%–90.3% in VMAT plans which produced the highest minimum doses, except for 1 patient (#1). Both D98 and D2 were similar in IMRT and VMAT plans (Table 1).
Fig. 2 shows a comparative exemplary dose-volume-histogram (DVH) with PTV and mean brain dose in IMRT and VMAT plans of patient #3, indicating sufficient and comparable PTV dose coverage with all treatment planning techniques and superior brain doses in the VMAT plan. Fig. 3 highlights depth dose characteristics of the VMAT plan from the surface of the scalp of patient #1.
2. Dosimetric comparison of organ at risk involvement of IMRT and VMAT treatment plans
Table 2 shows the organ at risk doses for each plan of each individual patients. In all patients, VMAT plans produced considerable lower mean brain doses (range, 13.8 to 21.2 Gy) compared to non-coplanar (range, 18.7 to 24 Gy) and coplanar IMRT plans (range, 19.7 to 25.8 Gy). The non-coplanar IMRT technique was slightly superior in mean brain dose.
The mean dose at the brain stem was lowest in VMAT, followed by non-coplanar IMRT and coplanar IMRT in three patients. In 2 patients, brain stem dose was superior in non-coplanar IMRT (Table 2). Maximum dose to the optic chiasm was lower in non-coplanar (range, 3.8 to 12.5 Gy; mean, 7.7 Gy) compared to coplanar (range, 3.8 to 17.6; mean, 11.1 Gy) IMRT in each of the 5 patient planning cases, while VMAT ranged intermediate (range, 4.7 to 13.3 Gy; mean, 8.4 Gy). In 4 patients, VMAT was inferior (range, 4.7 to 13.3 Gy) to non-coplanar IMRT (range, 3.8 to 12.5 Gy) in regard of mean brain stem dose and in 1 patient, VMAT was superior (5.3 vs. 6.3 Gy) (Table 2).
VMAT plans produced considerably lower doses at the optic nerves (8.2–9.9 Gy) compared to coplanar (8.7–18.5 Gy) and non-coplanar (10.7–21.9 Gy) which had the highest optic nerves dose in all patients. For eye lenses, VMAT was superior in 3 patient cases compared to non-coplanar and coplanar IMRT, however, doses to the lenses overall were comparable (Table 2).
Mean brain dose in all patients was lowest in VMAT (18.6 Gy), followed by non-coplanar (21.9 Gy) and coplanar IMRT (22.8 Gy) which produced similar mean brain doses. Mean brain stem dose was 16.5, 13.4, and 13.2 Gy in coplanar, non-coplanar IMRT and VMAT. Mean dose at optical structures including optical nerves left and right and optic chiasm was 9.1 Gy in VMAT, 14.5 Gy in non-coplanar and 11.5 Gy in coplanar IMRT. Fig. 2 presents an exemplary comparative DVH comparing maximum doses to the optical system in VMAT and IMRT treatment plans of patient #3, showing superior dose involvement of optical structures in the VMAT plan. In summa, doses to the optical system were 11.5, 14.5, and 9.1 Gy for coplanar, non-coplanar IMRT and VMAT.
Doses to the lenses were lowest in VMAT (5.2 Gy) and highest in non-coplanar IMRT (6.3 Gy); they were 5.9 Gy in coplanar IMRT plans.
Discussion and Conclusion
The treatment of extensive malignant scalp lesions remains a rare yet challenging and complex indication for radiotherapy. TSI, i.e. total scalp irradiation constitutes an adequate treatment for angiosarcoma, mycosis fungoides and extensive squamous cell carcinoma of the scalp [2,6,10]. Most commonly, conventional lateral electron-photon techniques [9,21], IMRT [10] but also tomotherapy [5], RapidArc and –in some cases–HDR brachytherapy is used [22]. However, clinical presentation, the individual shape of the scalp and dose variability make it impossible to define the single best treatment for all patients and often dictate the chosen technique.
We previously reported a comprehensive dosimetric comparison of coplanar and non-coplanar photon IMRT in the treatment of extensive scalp lesions and found that non-coplanar beams increased dose homogeneity and PTV coverage and reduced doses to OARs, particularly at the optic chiasm [12].
This treatment planning study describes our institution’s experience in TSI and comparatively evaluates coplanar and non-coplanar IMRT against VMAT.
Keeping treatment delivery times low, generally has the benefit of reduced intrafraction patient motion and lower patient exposure to leakage radiation. Recently, Inoue et al. [23] reported shorter TSI delivery time and reduced MU using a multi jaw-size concave arc technique. In our study which was based on a dose description of 50 Gy to the PTV in single fractions of 2 Gy, MUs per single fraction were lowest in VMAT (950.8) and did not differ considerably from those of non-coplanar (662.8) but were significantly lower than MUs in coplanar IMRT (974.2). In the literature, MUs for VMAT in TSI are reported ranging from 1,791 (60 Gy in single fractions of 2 Gy, 9 arcs) to 1,188 (20 Gy in single fractions of 2 Gy, 6 arcs) [24,25].
In this work, homogeneity and target volume coverage was slightly superior in VMAT, however both PTV coverage and HI were comparable between plans. Non-coplanar IMRT plans were most conformal and dose hotspots within and outside the target volume were comparable between treatment plans despite better maximum and minimum PTV doses VMAT plans. Hu et al. [25] showed improved target volume coverage and homogeneity for both IMRT and RapidArc and Song et al. [24] reported the best conformity in tomotherapy plans (CI, 1.35) followed by VMAT (CI, 1.49) while the photon-electron technique was least conformal (CI, 1.73). Our own results indicate comparable conformity between VMAT (CI, 1.43) and non-coplanar IMRT (CI, 1.45), while conformity was best with the coplanar IMRT technique in our study (CI, 1.4), however differences in conformity were smaller in our study compared to the data of Song et al. [24]. Analogue to the latter work, the HI in our study was similar across plans, yet, slightly poorer in IMRT plans compared to VMAT.
Overall, critical structures including brain, brain stem, optical system, and lenses could be spared best with VMAT which also produced the sharpest dose gradient towards healthy tissue but all critical organ doses were within clinically acceptable levels. In all patient planning cases, the optic structures received a mean dose less than 25 Gy which was also not exceeded for the brain in all except one coplanar IMRT plan. Mean doses to the lenses ranged from 4.5–8.1 Gy and never exceeded 7 Gy in VMAT and coplanar IMRT plans while the highest doses to the lenses were noted in a non-coplanar IMRT.
In the work of Hu et al., IMRT was inferior to 3D-conformal radiation therapy (CRT) in sparing of optical structures and RapidArc failed to improve doses to the optic system [25]. In our study, doses at the optic nerves were superior for VMAT in all patients compared to coplanar and non-coplanar IMRT while the optic chiasm could be spared best with non-coplanar IMRT. According to Hu et al. [25], doses to the lenses could be reduced by using RapidArc. VMAT in our study failed to reduce doses to the lenses compared to coplanar IMRT, however, sparing of the lenses was inferior in non-coplanar IMRT as opposed to VMAT. Song et al. [24] reported the lateral electronphoton technique to feature lower doses to optical structures compared to VMAT and helical tomotherapy however, irradiated brain volume and doses to the brain were superior in VMAT which is in accordance with our results. Compared to the results of Song et al. [24], optic nerves and brain stem doses were higher in our VMAT plans while doses to the lenses were comparable. The data of Hu et al. [25] indicates that 3D-CRT is clearly inferior to non-coplanar IMRT in regard of target coverage and brain protection where IMRT significantly deceased doses to the brain compared to 3D-CRT. In the work of Song et al. [24], brain protection was superior in VMAT and tomotherapy compared to the photon-electron approach which are further complicated by technical obstacles such as field matching.
Our own results suggest that both brain and brain stem protection was best in VMAT while the highest brain and brain stem doses were produced with the coplanar-IMRT plans.
In a clinical perspective, protection of the brain is prudent in the context of maintaining neurocognitive function in patients. In all treatment plans in our study, a mean brain dose of 25 Gy was not exceeded, however, recently there is increasing evidence that even low doses to the brain can significantly impact neurocognition [26] so that particular importance comes to protection of the hippocampus [27] which in our work could not be achieved due to the fact that treatment planning was CT-based and did not include cranial magnetic resonance imaging (MRI). MRI-based delineation, however, can help to further reduce the dose in the hippocampal region in order to prevent neurocognitive dysfunction when using VMAT for TSI. Song et al. [24] were able to show that using the lateral photon-electron or tomotherapy approach, doses to the hippocampus could be lowered compared to VMAT, however, not to the point of effective sparing of the hippocampal region.
In summa, the results of our study suggest superior target volume coverage, homogeneity, dose gradient and organ at risk protection in VMAT which provided particularly low doses to the brain and optical system. However, the variation of dosimetric parameters between patients and the small sample size in this study need to be considered in the interpretation of the results.
In the real clinical situation, additional limiting factors such as planning time, equipment availability but also the specific individual clinical presentation and both the institutes’ experience and technical capacities must be considered when choosing an adequate technique for TSI.