Dosimetric and LETd comparison of horizontal port and gantry carbon-ion radiotherapy for early-stage non–small cell lung cancer: an exploratory study

Article information

Radiat Oncol J. 2026;.roj.2026.00080
Publication date (electronic) : 2026 August 24
doi : https://doi.org/10.3857/roj.2026.00080
1Department of Radiation Oncology, Yonsei Cancer Center, Seoul, Republic of Korea
2Department of Radiation Oncology, Mayo Clinic Florida, Jacksonville, FL, USA
3Department of Radiation Oncology, Yonsei Cancer Center, Heavy Ion Therapy Research Institute, Yonsei University College of Medicine, Seoul, Republic of Korea
Correspondence: Chae-Seon Hong Department of Radiation Oncology, Yonsei Cancer Center, Heavy Ion Therapy Research Institute, Yonsei University College of Medicine, 50-1 Yonsei-ro, Seodaemun-gu, Seoul 03722, Korea E-mail: CS.HONG@yuhs.ac
Kyung Hwan Kim Department of Radiation Oncology, Yonsei Cancer Center, Heavy Ion Therapy Research Institute, Yonsei University College of Medicine, 50-1 Yonsei-ro, Seodaemun-gu, Seoul 03722, Korea E-mail: KYUNGHKIM@yuhs.ac
*These authors contributed equally to this work.
Received 2026 January 15; Revised 2026 July 3; Accepted 2026 July 6.

Abstract

Purpose

We aimed to evaluate the feasibility of using a horizontal port (HP) system for carbon ion radiotherapy (CIRT) in non–small cell lung cancer (NSCLC) by comparing the dosimetric characteristics and dose-averaged linear energy transfer (LETd) with those of a gantry-based plan.

Materials and Methods

Five patients with NSCLC treated with CIRT were included in this study. Both gantry-based and HP plans were generated using a single-field uniform dose optimization in RayStation 2024A with a pencil beam algorithm. Gantry plan beam angles followed institutional clinical protocols, whereas HP plans used beam angles of 90° or 270° with a couch roll of ±15° and yaw of ±30°. Plan evaluation was conducted using dose–volume histograms, LETd–volume histograms, and beam range analysis.

Results

Dose distributions for the clinical target volume (CTV) and ipsilateral lung were similar between plans. The CTV V95% was 100% in both plans. For the ipsilateral lung, mean V5 Gy (relative biological effectiveness [RBE]) and V20 Gy (RBE) were slightly lower in the gantry plan (12.40% and 5.77%, respectively, compared to 13.54% and 6.15%, respectively). Mean LETd98% within the CTV was higher in the gantry plan (65.27 keV/μm compared to 57.07 keV/μm). Beam energy and beam range were consistently lower and shorter, respectively, in the gantry plan.

Conclusion

The preliminary dosimetric performance of the HP plan was generally comparable to that of the gantry plan. However, these findings should be interpreted with caution. The gantry system may provide greater reliability and clinical efficiency owing to the simpler workflow and reduced sensitivity to setup variability and organ motion.

Introduction

Radiotherapy, along with chemotherapy and surgery, is one of the principal treatment modalities for cancer. Since the discovery of X-rays in 1895, radiotherapy has undergone continuous advancement for over a century [1]. During the early to mid-20th century, kilovoltage energies were commonly employed; however, subsequent technological progress led to the adoption of megavoltage energies as the clinical standard. This transition enabled the development of a range of megavoltage-based treatment techniques, including three-dimensional conformal radiotherapy, intensity-modulated radiotherapy (IMRT), volumetric modulated arc therapy, and stereotactic body radiotherapy. Among these, IMRT is particularly well-suited for cases involving complex anatomy or tumors located adjacent to critical organs at risk (OARs), and it is therefore widely used in the treatment of head and neck, breast, prostate, and lung cancers [2-6].

In 1946, Wilson first proposed that the Bragg peak characteristic of protons could be exploited for cancer treatment [7]. In 1954, clinical studies were initiated at the University of California, Berkeley [8]. In these early investigations, proton beams were not applied to malignant tumors; instead, irradiation of the pituitary gland was performed to treat disorders associated with abnormal hormone production [9,10]. At present, proton therapy is commonly used in the management of various malignancies, including head and neck, prostate, lung, and liver cancers [11-15].

Carbon ions, similar to protons, can deliver highly focused radiation to the target by exploiting the Bragg peak, and they exhibit less lateral scattering than protons [16]. In addition, at the Bragg peak, carbon ions demonstrate a higher relative biological effectiveness (RBE), approximately 2–3 times that of protons, along with a higher linear energy transfer (LET). Owing to these advantages, carbon ion radiotherapy (CIRT) has emerged as one of the most promising radiotherapy modalities worldwide, and it is particularly attractive for treating radioresistant malignancies such as sarcomas, melanomas, and certain adenocarcinomas [17-19]. The National Institute of Radiological Sciences in Japan, which initiated carbon ion therapy in 1994, is among the few institutions with over 30 years of experience in CIRT, and it consistently reports clinical outcomes [20-22]. Additional CIRT facilities have now been established in Gunma, Osaka, and Yamagata in Japan [23-25], as well as in Heidelberg, Germany [26]. In South Korea, the Yonsei Cancer Center began CIRT in 2023, becoming the first facility in the country to offer this treatment [27]. In addition, the Mayo Clinic Florida is currently constructing the first CIRT facility in the United States [28].

CIRT equipment can be broadly categorized into fixed-beam port and gantry systems, each with distinct advantages and limitations. Fixed-beam port systems use horizontal and/or vertical beam ports and are associated with lower installation costs. In contrast, gantry systems enable irradiation from patient-specific optimal beam angles, providing greater flexibility in treatment delivery; however, they require higher installation costs and involve more complex mechanical structures. Accordingly, gantry systems are currently available in only four of the 17 CIRT facilities worldwide, and in only five of the 53 treatment rooms [29].

In lung cancer treatment, CIRT can reduce the volume of low-dose irradiation to normal lung tissue owing to its small lateral penumbra compared with other treatment modalities. In studies of early-stage non–small cell lung cancer (NSCLC), Miyamoto et al. [30,31] reported that CIRT achieved outcomes comparable to those of surgery for stage I disease. Demizu et al. [32] further demonstrated that CIRT provides a superior dose distribution compared with stereotactic body radiotherapy, thereby offering a safer option for patients with impaired lung function. Similarly, Okano et al. [33] reported excellent local control with CIRT and no significant adverse effects in patients with early-stage lung cancer and interstitial lung disease. For locally advanced NSCLC, Hayashi et al. [34] suggested that CIRT may represent an effective treatment alternative for patients who are unable to undergo surgery or chemotherapy, with acceptable toxicity profiles.

In CIRT, beam angle selection is an important consideration in treatment planning because it affects target dose conformity, normal tissue sparing, and the spatial distribution of the high-LET region near the distal end of the beam. Although gantry systems offer greater flexibility, fixed-beam port systems remain more widely used due to cost and facility-related limitations. Several studies have therefore compared the feasibility of fixed-beam port systems with that of gantry systems. Koom et al. [35] showed that, in pancreatic cancer, a gantry system achieved equivalent target coverage while reducing dose to surrounding organs compared with a fixed-beam port system. Miyasaka et al. [36] compared a fixed-beam port system with couch roll to a gantry system in liver cancer and reported no significant differences in doses to OARs, except for the clinical target volume (CTV) and gross tumor volume. Yagi et al. [37] evaluated nine cancer sites, including lung cancer, and found that a horizontal port (HP) system with couch roll was effective for centrally located tumors, whereas its effectiveness varied according to tumor location. These findings suggest that the feasibility of HP-based CIRT for lung cancer remains uncertain because of tumor location, thoracic anatomy, and tissue heterogeneity. Moreover, although dose-averaged LET (LETd) is increasingly recognized as an important metric in CIRT, LETd-based comparisons between HP and gantry systems for lung cancer remain limited. Therefore, this exploratory study evaluated the feasibility of HP-based CIRT for early-stage NSCLC by comparing RBE-weighted dose and LETd metrics with those of gantry-based plans.

Materials and Methods

1. Patient selection and image acquisition

Five patients who underwent CIRT for lung cancer between June 2024 and January 2025 were included in this study, and their clinical characteristics are summarized in Table 1. All patients received respiratory training before undergoing a four-dimensional computed tomography (4D CT) scan. The scan was then acquired under free-breathing conditions using a 4D CT scanner (Aquilion ONE Vision, Canon Medical Systems, Otawara, Japan), with respiratory signals recorded using an Anzai respiratory monitoring system (AZ-733V, Anzai Medical, Tokyo, Japan).

Patient characteristics

2. Treatment planning

The CTV was delineated by a radiation oncologist in each phase of the 4D CT images. Both gantry and HP plans were generated using RayStation 2024A (RaySearch Laboratories, Stockholm, Sweden) with the pencil-beam dose-calculation algorithm. RBE-weighted doses were calculated using the microdosimetric kinetic model [38]. The prescribed dose was 60 Gy (RBE) to the CTV, delivered in four fractions. Dose constraints for the CTV and OARs were as follows: CTV V95% (volume receiving at least 95% of the prescribed dose) > 99.9%; ipsilateral lung V5 Gy (RBE) (volume receiving at least 5 Gy [RBE]) < 55% and V20 Gy (RBE) < 25%; spinal cord Dmax (maximum dose within the organ) < 15 Gy (RBE) and V16 Gy (RBE) < 0.2 cm³; bronchus V26 Gy (RBE) < 0.2 cm³; trachea V30 Gy (RBE) < 0.2 cm³; and esophagus V20 Gy (RBE) < 0.2 cm³. Treatment planning was performed using a single-field uniform dose optimization approach, with each fraction delivering 15 Gy (RBE). Fig. 1 shows the couch coordinate system and the setup configurations for the gantry and HP plans. As this was a retrospective study, the HP plans were generated in the gantry system using the available computed tomography (CT) images. The intended HP configuration consisted of horizontal beam directions, corresponding to gantry angles of 90° or 270°, with a ±15° couch roll and a ±30° couch yaw. Since the required couch roll exceeded the allowable range in the treatment planning system, it was represented by modifying the gantry angle. Accordingly, HP configurations with gantry angles of 90° and 270° combined with a ±15° couch roll were implemented using gantry angles of 75°/105° and 255°/285°, respectively (Table 2). For four-dimensional robust optimization, setup and range uncertainties were set to 3 mm and 3.5%, respectively. The gating phase was selected from phases in which the displacement of the CTV isocenter is within 2 mm relative to the end-of-exhale phase (Table 3).

Fig. 1.

(A) Definition of couch rotation directions (yaw, pitch, and roll). Yaw is the rotation around the posterior–anterior axis, pitch is the rotation around the right–left axis, and roll is the rotation around the inferior–superior axis. For a head-first supine patient, positive angles are defined as clockwise rotations when viewed from below, right, and feet for yaw, pitch, and roll, respectively. (B, C) The beam configurations for patient A in the treatment planning system, corresponding to the gantry system (B) and the horizontal port (HP) system emulated using the gantry system (C), respectively. In this figure, G, Y, and R denote the gantry angle, couch yaw, and couch roll, respectively. Parentheses (≈) represent the intended beam configurations for the HP system.

Gantry and couch angle information for each patient

Treatment planning and computation settings

3. Dosimetric, LETd, and beam range evaluation

For dosimetric evaluation, HP plans were compared with gantry plans using dose–volume histogram (DVH) parameters. We evaluated the following DVH parameters: V95% for CTV; V5 Gy (RBE) and V20 Gy (RBE) for the ipsilateral lung; Dmax and V16 Gy (RBE) for the spinal cord; V26 Gy (RBE) for the bronchus; V30 Gy (RBE) for the trachea; and V20 Gy (RBE) for the esophagus. In addition, conformity index (CI) and homogeneity index (HI) were evaluated to assess dose conformity and homogeneity within the target volume. The Paddick CI and HI were calculated using equations (1) and (2), respectively [39-41],

(1) CI=TVPIV2TV×PIV,

and

(2) HI=D2%−D98%D50%,

where TV denotes the target volume, PIV the volume encompassed by the 95% prescription isodose, and TVPIV the portion of the target volume covered by the PIV. A CI value closer to 1 indicates better dose conformity, whereas an HI value closer to 0 reflects better dose homogeneity.

For LETd evaluation, LETd–volume histogram parameters were analyzed, including V40 keV/μm and LETd98% for the CTV, as proposed by Li et al. [42].

Finally, because the impact of range uncertainty depends on the absolute beam range, beam ranges were compared between the gantry and HP plans. Beam ranges were estimated using an energy–range table in water measured at our institution.

Results

1. Dosimetric evaluation

Fig. 2 shows the mean DVHs for the CTV and OARs of the five patients in the gantry and HP plans. The evaluated DVH parameters, including the CI and HI, are summarized as mean ± standard deviation in Table 4, and the individual patient values are provided in Supplementary Table S1. CTV V95% was 100% in all plans, satisfying the institutional dose constraints. The ipsilateral lung V5 Gy (RBE) was 12.4% ± 2.77% for the gantry plans and 13.54% ± 1.19% for the HP plans, while the corresponding V20 Gy (RBE) values were 5.77% ± 0.75% and 6.15% ± 0.30%, respectively. The maximum dose to the spinal cord was 1.07 ± 1.92 Gy (RBE) for the gantry plans and 1.87 ± 1.51 Gy (RBE) for the HP plans. The V16 Gy (RBE) of the spinal cord, V30 Gy (RBE) of the trachea, and V20 Gy (RBE) of the esophagus were 0.0 cm³ in all cases. The bronchus V26 Gy (RBE) was 0.09 cm³ only in the HP plan of patient C and 0.0 cm³ in all other cases. The CI was 0.34 ± 0.07 for the gantry plans and 0.32 ± 0.07 for the HP plans, while the HI was 0.02 ± 0.01 and 0.02 ± 0.00, respectively.

Fig. 2.

Mean dose–volume histograms for the clinical target volume (CTV) and organs at risk in the gantry and horizontal port (HP) plans for the five patients. RBE, relative biological effectiveness.

Dose-volume histogram parameters of the gantry and HP plans

2. LETd evaluation

Mean LETd–volume histograms showed that the minimum LETd within the CTV was higher in the gantry plan than in the HP plans (Fig. 3). Mean V40 keV/μm and LETd98% values for the CTV are presented in Table 5, with values for each patient in Supplementary Table S2. V40 keV/μm for the CTV was 100% in all plans. Mean LETd98% of the CTV was 65.27 keV/μm for the gantry plans and 57.07 keV/μm for the HP plans, with the gantry plans showing a higher value by 8.2 keV/μm.

Fig. 3.

Mean dose-averaged linear energy transfer (LETd)–volume histograms for the gantry and horizontal port (HP) plans in five patients. The red vertical line indicates the LETd threshold of 36.4 keV/μm reported by Li et al. [42] as a reference value related to local recurrence in non–small cell lung cancer. CTV, clinical target volume.

LETd-volume histogram parameters of the gantry and HP plans for five patients

3. Beam range in water

Fig. 4 presents the beam energy and beam range in water for the five patients. In all patients, the beam energy and beam range were lower and shorter, respectively, under the gantry plan than under the HP plan.

Fig. 4.

Box plots of beam energy and corresponding range in water for the gantry and horizontal port (HP) plans in each patient. The left y-axis represents beam energy (MeV/u), and the right y-axis represents range in water (g/cm²).

Discussion and Conclusion

This study evaluated the clinical feasibility of using the HP system in CIRT for five patients with NSCLC. In the dosimetric evaluation, the CTV coverage achieved with the HP plan was comparable to that of the gantry plan, and doses to OARs satisfied institutional constraints. One limitation of the HP system is the restricted beam angle selection, which may lead to increased OAR doses in certain scenarios; however, in this study, adequate separation between the target and adjacent OARs allowed all constraints to be met. Consistent with these findings, Yagi et al. [37] reported no significant differences between gantry and HP plans. In lung tissue, the relatively low stopping power ratio can result in dose extension beyond the distal edge of the target. As illustrated in Fig. 5, the low-dose region extended predominantly in the anterior–posterior direction in the gantry plan, whereas extension occurred mainly in the superior–inferior direction in the HP plan.

Fig. 5.

Representative dose distributions of the gantry and horizontal port plans for patient A. RBE, relative biological effectiveness.

For LETd evaluation, both the gantry and HP plans achieved a V40 keV/µm of 100%, with mean LETd98% values of 65.27 keV/µm and 57.07 keV/µm, respectively. Li et al. [42] reported that, among 62 patients with locally advanced NSCLC, 16 experienced local recurrence. In that study, the mean LETd98% was 37.5 keV/µm in the local control group and 32.0 keV/µm in the local recurrence group, while the corresponding mean V40 keV/µm values were 89.2% and 75.4%. Based on these results, Li et al. [42] proposed that the LETd98% of the internal gross tumor volume should exceed 36.4 keV/μm and that V40 keV/µm should exceed 88%. In the present study, both planning approaches satisfied these LETd criteria. For the ipsilateral lung, LETd-based evaluation metrics have not yet been established. Although relatively high LETd values were observed in the ipsilateral lung, these regions were confined to low-dose areas, and their clinical relevance is therefore likely to be limited.

Beam ranges in water differed between the gantry and HP plans due to differences in beam angle and tissue heterogeneity along the beam paths. The beam ranges in the gantry plan were shorter than those in the HP plan for all patients. Although the differences in beam range may be associated with the LETd differences observed within the CTV between the two plans, LETd is influenced by multiple factors, such as beam arrangement, tissue heterogeneity, beam modulation, and optimization strategy. Therefore, these factors should be considered collectively when evaluating LETd.

This study has several limitations. First, the cohort was small, comprising only five patients; therefore, statistical significance was not evaluated, which limits the generalizability of the findings. Second, implementation of the HP system requires CT scans to be acquired with couch roll. However, additional CT scans with couch roll were not acquired because this was a retrospective study. Therefore, the HP configuration was emulated in the gantry system by adjusting the gantry angle instead of applying couch roll. Third, the potential uncertainties associated with couch roll were not evaluated in this study. When couch roll is performed, internal organs and soft tissues may shift or deform due to gravitational effects [43]. Such anatomical changes may affect the beam range [44]; however, these uncertainties were not considered in this study. Finally, the pencil beam algorithm used in this study may introduce dose calculation uncertainties in heterogeneous regions, such as the lung, compared with Monte Carlo calculations [45-47].

In conclusion, within this small exploratory feasibility study, the HP plan showed preliminary dosimetric findings that were generally comparable to those of the gantry plan. However, given the small cohort, these findings should be interpreted with caution and not be overgeneralized. When non-planning factors such as setup stability, organ motion, and reproducibility are considered, gantry-based delivery may offer greater clinical reliability. In addition, HP-based treatment would require two CT simulations and deformable image registration, whereas the gantry system eliminates these steps, potentially reducing clinical workload and improving overall treatment efficiency.

Notes

Statement of Ethics

The protocol of this study was approved by the Institutional Review Board of Yonsei University College of Medicine, Severance Hospital (IRB No. 4-2025-0209). The requirement for informed consent was waived by the Institutional Review Board owing to the retrospective nature of the study.

Conflict of Interest

No potential conflict of interest relevant to this article was reported.

Funding

This work was supported by the National Research Foundation of Korea (NRF) grant funded by the Korea government (MSIT) (RS-2024-00455777).

Author Contributions

Conceptualization, KMF, SHB, KHK, CSH; Methodology, SHB, KMF, CSH; Investigation, SHB, KMF, SP, JCP, MCH, CK, JSK, CJB, KHK, CSH; Formal analysis, SHB, KMF; Resources, SP, JCP, MCH, CK, JSK, KHK, CSH; Project administration, KMF, KHK, CSH; Writing of original draft, SHB, KMF, CSH; Writing of the review and editing, all authors; Approval of final manuscript, all authors.

Data Availability Statement

The data supporting the conclusions of this study can be obtained from the corresponding author upon a reasonable request.

Supplementary Materials

Supplementary materials can be found via https://doi.org/10.3857/roj.2026.00080.

Supplementary Table S1.

Dose-volume histogram parameters of the gantry and HP plans for individual patients

roj-2026-00080-Supplementary-Table-S1.pdf
Supplementary Table S2.

LETd-volume histogram parameters of the gantry and HP plans for individual patients

roj-2026-00080-Supplementary-Table-S2.pdf

References

1. Thariat J, Hannoun-Levi JM, Sun Myint A, Vuong T, Gerard JP. Past, present, and future of radiotherapy for the benefit of patients. Nat Rev Clin Oncol 2013;10:52–60. 10.1038/nrclinonc.2012.203. 23183635.
2. Garden AS, Morrison WH, Rosenthal DI, Chao KS, Ang KK. Target coverage for head and neck cancers treated with IMRT: review of clinical experiences. Semin Radiat Oncol 2004;14:103–9. 10.1053/j.semradonc.2003.12.004. 15095256.
3. McDonald MW, Godette KD, Butker EK, Davis LW, Johnstone PA. Long-term outcomes of IMRT for breast cancer: a single-institution cohort analysis. Int J Radiat Oncol Biol Phys 2008;72:1031–40. 10.1016/j.ijrobp.2008.02.053. 18440727.
4. Chang JS, Chang JH, Kim N, Kim YB, Shin KH, Kim K. Intensity modulated radiotherapy and volumetric modulated arc therapy in the treatment of breast cancer: an updated review. J Breast Cancer 2022;25:349–65. 10.4048/jbc.2022.25.e37. 36265885.
5. Fischer-Valuck BW, Rao YJ, Michalski JM. Intensity-modulated radiotherapy for prostate cancer. Transl Androl Urol 2018;7:297–307. 10.21037/tau.2017.12.16. 30050791.
6. Chan C, Lang S, Rowbottom C, Guckenberger M, Faivre-Finn C, ; IASLC Advanced Radiation Technology Committee. Intensity-modulated radiotherapy for lung cancer: current status and future developments. J Thorac Oncol 2014;9:1598–608. 10.1097/jto.0000000000000346. 25436795.
7. Wilson RR. Radiological use of fast protons. Radiology 1946;47:487–91. 10.1148/47.5.487. 20274616.
8. Boone ML, Lawrence JH, Connor WG, Morgado R, Hicks JA, Brown RC. Introduction to the use of protons and heavy ions in radiation therapy: historical perspective. Int J Radiat Oncol Biol Phys 1977;3:65–9. 10.1016/0360-3016(77)90229-2. 96079.
9. Lawrence JH, Tobias CA, Born JL, et al. Pituitary irradiation with high-energy proton beams: a preliminary report. Cancer Res 1958;18:121–34. 13511365.
10. Endo M. Robert R. Wilson (1914-2000): the first scientist to propose particle therapy: use of particle beam for cancer treatment. Radiol Phys Technol 2018;11:1–6. 10.1007/s12194-017-0428-z. 29058267.
11. Tokuuye K, Akine Y, Kagei K, et al. Proton therapy for head and neck malignancies at Tsukuba. Strahlenther Onkol 2004;180:96–101. 10.1007/s00066-004-1132-6. 14762662.
12. Bush DA, Slater JD, Bonnet R, et al. Proton-beam radiotherapy for early-stage lung cancer. Chest 1999;116:1313–9. 10.1378/chest.116.5.1313. 10559093.
13. Shioyama Y, Tokuuye K, Okumura T, et al. Clinical evaluation of proton radiotherapy for non-small-cell lung cancer. Int J Radiat Oncol Biol Phys 2003;56:7–13. 10.1016/s0360-3016(02)04416-4. 12694818.
14. Chuong M, Kaiser A, Molitoris J, Mendez Romero A, Apisarnthanarax S. Proton beam therapy for liver cancers. J Gastrointest Oncol 2020;11:157–65. 10.21037/jgo.2019.04.02. 32175119.
15. Brada M, Pijls-Johannesma M, De Ruysscher D. Proton therapy in clinical practice: current clinical evidence. J Clin Oncol 2007;25:965–70. 10.1200/jco.2006.10.0131. 17350945.
16. Suit H, DeLaney T, Goldberg S, et al. Proton vs carbon ion beams in the definitive radiation treatment of cancer patients. Radiother Oncol 2010;95:3–22. 10.1016/j.radonc.2010.01.015. 20185186.
17. Tinganelli W, Durante M. Carbon ion radiobiology. Cancers (Basel) 2020;12:3022. 10.3390/cancers12103022. 33080914.
18. Byun HK, Han MC, Yang K, et al. Physical and biological characteristics of particle therapy for oncologists. Cancer Res Treat 2021;53:611–20. 10.4143/crt.2021.066. 34139805.
19. Kamada T. Clinical evidence of particle beam therapy (carbon). Int J Clin Oncol 2012;17:85–8. 10.1007/s10147-012-0388-6. 22426888.
20. Tsujii H, Mizoe J, Kamada T, et al. Clinical results of carbon ion radiotherapy at NIRS. J Radiat Res 2007;48 Suppl A:A1–13. 10.1269/jrr.48.a1. 17513896.
21. Kamada T, Tsujii H, Blakely EA, et al. Carbon ion radiotherapy in Japan: an assessment of 20 years of clinical experience. Lancet Oncol 2015;16:e93–100. 10.1016/s1470-2045(14)70412-7. 25638685.
22. Mohamad O, Makishima H, Kamada T. Evolution of carbon ion radiotherapy at the National Institute of Radiological Sciences in Japan. Cancers (Basel) 2018;10:66. 10.3390/cancers10030066. 29509684.
23. Ohno T, Kanai T, Yamada S, et al. Carbon ion radiotherapy at the Gunma University Heavy Ion Medical Center: new facility set-up. Cancers (Basel) 2011;3:4046–60. 10.3390/cancers3044046. 24213124.
24. Hamatani N, Tsubouchi T, Takashina M, Yagi M, Kanai T. Commissioning of carbon-ion radiotherapy for moving targets at the Osaka Heavy-Ion Therapy Center. Med Phys 2022;49:801–12. 10.1002/mp.15403. 34894413.
25. Souda H, Kanai T, Lee SH, et al. Commissioning and first operation of East Japan Heavy Ion Center at Yamagata University. J Phys Conf Ser 2024;2743:012092. 10.1088/1742-6596/2743/1/012092.
26. Rieken S, Habermehl D, Haberer T, Jaekel O, Debus J, Combs SE. Proton and carbon ion radiotherapy for primary brain tumors delivered with active raster scanning at the Heidelberg Ion Therapy Center (HIT): early treatment results and study concepts. Radiat Oncol 2012;7:41. 10.1186/1748-717x-7-41. 22436135.
27. Han MC, Choi SH, Hong CS, et al. The first Korean carbon-ion radiation therapy facility: current status of the Heavy-ion Therapy Center at the Yonsei Cancer Center. Radiat Oncol J 2024;42:295–307. 10.3857/roj.2024.00206. 39748530.
28. Beltran C, Amos RA, Rong Y. We are ready for clinical implementation of Carbon Ion Radiotherapy in the United States. J Appl Clin Med Phys 2020;21:6–9. 10.1002/acm2.13133.
29. Particle Therapy Co-Operative Group. Facilities in operation [Internet]. Geneva: Particle Therapy Co-Operative Group [cited 2026 May 31]. Available from: https://www.ptcog.site/index.php/facilities-in-operation-public.
30. Miyamoto T, Yamamoto N, Nishimura H, et al. Carbon ion radiotherapy for stage I non-small cell lung cancer. Radiother Oncol 2003;66:127–40. 10.1016/s0167-8140(02)00367-5. 12648784.
31. Miyamoto T, Baba M, Sugane T, et al. Carbon ion radiotherapy for stage I non-small cell lung cancer using a regimen of four fractions during 1 week. J Thorac Oncol 2007;2:916–26. 10.1097/jto.0b013e3181560a68. 17909354.
32. Demizu Y, Fujii O, Iwata H, Fuwa N. Carbon ion therapy for early-stage non-small-cell lung cancer. Biomed Res Int 2014;2014:727962. 10.1155/2014/727962. 25295269.
33. Okano N, Suefuji H, Nakajima M, et al. Clinical results of carbon-ion radiotherapy for stage I non-small cell lung cancer with concomitant interstitial lung disease: a Japanese national registry study (J-CROS-LUNG). J Radiat Res 2023;64(Suppl 1):i2–7. 10.1093/jrr/rrad008. 37036751.
34. Hayashi K, Yamamoto N, Nakajima M, et al. Clinical outcomes of carbon-ion radiotherapy for locally advanced non-small-cell lung cancer. Cancer Sci 2019;110:734–41. 10.1111/cas.13890. 30467928.
35. Koom WS, Mori S, Furuich W, Yamada S. Beam direction arrangement using a superconducting rotating gantry in carbon ion treatment for pancreatic cancer. Br J Radiol 2019;92:20190101. 10.1259/bjr.20190101. 30943057.
36. Miyasaka Y, Lee SH, Souda H, et al. Treatment planning comparison of gantry-based and fixed beams for the treatment of liver tumors with carbon ion therapy. In Vivo 2024;38:3002–10. 10.21873/invivo.13783. 39477437.
37. Yagi M, Beltran CJ, Shimizu S, et al. Carbon ion therapy for laterally located tumors require multiple fixed ports or a rotating gantry. Med Dosim 2024;49:279–90. 10.1016/j.meddos.2024.02.003. 38824052.
38. Wang W, Sun W, Shen H, Zhao J. Validation of the relative biological effectiveness of active-energy scanning carbon-ion radiotherapy on a commercial treatment planning system with a microdosimetic kinetic model. Radiat Oncol 2023;18:82. 10.1186/s13014-023-02267-8. 37198685.
39. Paddick I. A simple scoring ratio to index the conformity of radiosurgical treatment plans: technical note. J Neurosurg 2000;93 Suppl 3:219–22. 10.3171/jns.2000.93.supplement. 11143252.
40. International Commission on Radiation Units and Measurements. Prescribing, recording, and reporting photon-beam intensity-modulated radiation therapy (IMRT). J Int Comm Radiat Units Meas 2010;10:1–106. 10.1093/jicru_ndq002.
41. Hong CS, Ju SG, Ahn YC, et al. Normal lung sparing tomotherapy technique in stage III lung cancer. Radiat Oncol 2017;12:167. 10.1186/s13014-017-0905-x. 29110732.
42. Li G, Ma N, Wang W, et al. Dose-averaged linear energy transfer within the gross tumor volume of non-small-cell lung cancer affects the local control in carbon-ion radiotherapy. Radiother Oncol 2024;201:110584. 10.1016/j.radonc.2024.110584.
43. Buckley JG, Rai R, Liney GP, et al. Anatomical deformation due to horizontal rotation: towards gantry-free radiation therapy. Phys Med Biol 2019;64:175014. 10.1088/1361-6560/ab324c. 31307023.
44. Liebl J, Paganetti H, Zhu M, Winey BA. The influence of patient positioning uncertainties in proton radiotherapy on proton range and dose distributions. Med Phys 2014;41:091711. 10.1118/1.4892601. 25186386.
45. Schreuder AN, Bridges DS, Rigsby L, et al. Validation of the RayStation Monte Carlo dose calculation algorithm using a realistic lung phantom. J Appl Clin Med Phys 2019;20:127–37. 10.1002/acm2.12777.
46. Maes D, Saini J, Zeng J, Rengan R, Wong T, Bowen SR. Advanced proton beam dosimetry part II: Monte Carlo vs. pencil beam-based planning for lung cancer. Transl Lung Cancer Res 2018;7:114–21. 10.21037/tlcr.2018.04.04. 29876310.
47. Ruangchan S, Palmans H, Knausl B, Georg D, Clausen M. Dose calculation accuracy in particle therapy: comparing carbon ions with protons. Med Phys 2021;48:7333–45. 10.1002/mp.15209. 34482555.

Article information Continued

Fig. 1.

(A) Definition of couch rotation directions (yaw, pitch, and roll). Yaw is the rotation around the posterior–anterior axis, pitch is the rotation around the right–left axis, and roll is the rotation around the inferior–superior axis. For a head-first supine patient, positive angles are defined as clockwise rotations when viewed from below, right, and feet for yaw, pitch, and roll, respectively. (B, C) The beam configurations for patient A in the treatment planning system, corresponding to the gantry system (B) and the horizontal port (HP) system emulated using the gantry system (C), respectively. In this figure, G, Y, and R denote the gantry angle, couch yaw, and couch roll, respectively. Parentheses (≈) represent the intended beam configurations for the HP system.

Fig. 2.

Mean dose–volume histograms for the clinical target volume (CTV) and organs at risk in the gantry and horizontal port (HP) plans for the five patients. RBE, relative biological effectiveness.

Fig. 3.

Mean dose-averaged linear energy transfer (LETd)–volume histograms for the gantry and horizontal port (HP) plans in five patients. The red vertical line indicates the LETd threshold of 36.4 keV/μm reported by Li et al. [42] as a reference value related to local recurrence in non–small cell lung cancer. CTV, clinical target volume.

Fig. 4.

Box plots of beam energy and corresponding range in water for the gantry and horizontal port (HP) plans in each patient. The left y-axis represents beam energy (MeV/u), and the right y-axis represents range in water (g/cm²).

Fig. 5.

Representative dose distributions of the gantry and horizontal port plans for patient A. RBE, relative biological effectiveness.

Table 1.

Patient characteristics

Patient CTV volume (cm3) Tumor location Patient position Prescription dose (Gy [RBE]) No. of fractions
A 4.43 Right middle lobe Prone 60 4
B 8.42 Right upper lobe Prone 60 4
C 5.73 Right middle lobe Supine 60 4
D 5.92 Left upper lobe Supine 60 4
E 8.27 Left lower lobe Prone 60 4

CTV, clinical target volume; RBE, relative biological effectiveness.

Table 2.

Gantry and couch angle information for each patient

Patient Port Gantry plan HP plan
Gantry angle (°) Couch yaw (°) Couch roll (°) Gantry anglea) (°) Couch yaw (°) Couch roll (°)
A, B, D 1 0 0 0 75 30 0
2 30 0 0 75 330 0
3 60 0 0 105 30 0
4 330 0 0 105 330 0
C 1 270 0 0 255 30 0
2 300 0 0 255 330 0
3 330 0 0 285 30 0
4 30 0 0 285 330 0
E 1 270 0 0 255 30 0
2 300 0 0 255 330 0
3 330 0 0 285 30 0
4 0 0 0 285 330 0

a)For the horizontal port (HP) plans, couch roll was represented by gantry angle modification using computed tomography images acquired without couch roll because of the retrospective study design. Thus, gantry angles of 75°/105° and 255°/285° represented HP configurations with horizontal beam directions of 90° and 270°, respectively, combined with a ±15° couch roll.

Table 3.

Treatment planning and computation settings

Plan Energy layer spacing 4D robust optimization Number of layers rescanning
Setup uncertainty Range uncertainty Organ motion uncertainty
Gantry Automatic (0.7) 3.0 mm 3.5% Gating windowa) 6
Horizontal port Automatic (0.7)

Energy layer spacing was set to Automatic, which uses variable spacing based on the Bragg peak width.

CT phase refers to the respiratory phase of four-dimensional computed tomography.

a)The gating phase was defined as the respiratory phase in which the clinical target volume isocenter displacement was within 2 mm relative to the end-of-exhalation position, corresponding to CT40–60% for patient A, CT40–80% for patient B, CT30–70% for patient C, and CT40–70% for patients D and E.

Table 4.

Dose-volume histogram parameters of the gantry and HP plans

Structure Parameter Gantry plan HP plan
CTV V95% (%) 100.00 ± 0.00 100.00 ± 0.00
CI 0.34 ± 0.07 0.32 ± 0.07
HI 0.02 ± 0.01 0.02 ± 0.00
Ipsilateral lung V5 Gy (RBE) (%) 12.40 ± 2.77 13.54 ± 1.19
V20 Gy (RBE) (%) 5.77 ± 0.75 6.15 ± 0.30
Spinal cord Dmax (Gy [RBE]) 1.07 ± 1.92 1.87 ± 1.51
V16 Gy (RBE) (cm³) 0.00 ± 0.00 0.00 ± 0.00
Trachea V30 Gy (RBE) (cm³) 0.00 ± 0.00 0.00 ± 0.00
Bronchus V26 Gy (RBE) (cm³) 0.00 ± 0.02 0.02 ± 0.04
Esophagus V20 Gy (RBE) (cm³) 0.00 ± 0.00 0.00 ± 0.00

Values are presented as mean±standard deviation.

HP, horizontal port; CTV, clinical target volume; V95%, volume of the clinical target volume receiving at least 95% of the prescribed dose; CI, conformity index; HI, homogeneity index; VX Gy, volume receiving at least X Gy (RBE); RBE, relative biological effectiveness; Dmax, maximum dose.

Table 5.

LETd-volume histogram parameters of the gantry and HP plans for five patients

Structure Parameter Gantry plan HP plan
CTV V40 keV/μm (%) 100.00 ± 0.00 100.00 ± 0.00
LETd98% (keV/μm) 65.27 ± 5.92 57.07 ± 5.96

Values are presented as mean±standard deviation.

LETd, dose-averaged linear energy transfer; HP, horizontal port; V40 keV/μm, percentage volume receiving at least 40 keV/μm LETd; LETd98%, LETd value covering 98% of the CTV.