J Korean Soc Ther Radiol Oncol > Volume 22(3); 2004 > Article
The Journal of the Korean Society for Therapeutic Radiology and Oncology 2004;22(3): 225-233.
Comparison of Virtual Wedge versus Physical Wedge Affecting on Dose Distribution of Treated Breast and Adjacent Normal Tissue for Tangential Breast Irradiation
Yeon Sil Kim, Sung Whan Kim, Sei Chul Yoon, Jung Seok Lee, Seok Hyun Son, Ihl Bong Choi
Department of Radiation Oncology, College of Medicine, The Catholic University of Korea, Seoul, Korea. yeonkim7@catholic.ac.kr
ABSTRACT
PURPOSE:
The ideal breast irradiation method should provide an optimal dose distribution in the treated breast volume and a minimum scatter dose to the nearby normal tissue. Physical wedges have been used to improve the dose distribution in the treated breast, but unfortunately introduce an increased scatter dose outside the treatment field, particularly to the contralateral breast. The typical physical wedge (PW) was compared with the virtual wedge (VW) to determine the difference in the dose distribution affecting on the treated breast and the contralateral breast, lung, heart and surrounding peripheral soft tissue. METHODS AND MATERIALS: The data collected consisted of a measurement taken with solid water, a Humanoid Alderson Rando phantom and patients. The radiation doses at the ipsilateral breast and skin, contralateral breast and skin, surrounding peripheral soft tissue, and ipsilateral lung and heart were compared using the physical wedge and virtual wedge and the radiation dose distribution and DVH of the treated breast were compared. The beam-on time of each treatment technique was also compared. Furthermore, the doses at treated breast skin, contralateral breast skin and skin 1.5 cm away from the field margin were also measured using TLD in 7 patients of tangential breast irradiation and compared the results with phantom measurements.
RESULTS:
The virtual wedge showed a decreased peripheral dose than those of a typical physical wedge at 15degrees, 30degrees, 45degrees, and 60degrees. According to the TLD measurements with 15degrees and 30degrees virtual wedge, the irradiation dose decreased by 1.35% and 2.55% in the contralateral breast and by 0.87% and 1.9% in the skin of the contralateral breast respectively. Furthermore, the irradiation dose decreased by 2.7% and 6.0% in the ipsilateral lung and by 0.96% and 2.5% in the heart. The VW fields had lower peripheral doses than those of the PW fields by 1.8% and 2.33%. However the skin dose increased by 2.4% and 4.58% in the ipsilateral breast. VW fields, in general, use less monitor units than PW fields and shortened beam-on time about half of PW. The DVH analysis showed that each delivery technique results in comparable dose distribution in treated breast.
CONCLUSIONS:
A modest dose reduction to the surrounding normal tissue and uniform target homogeneity were observed using the VW technique compare to the PW beam in tangential breast irradiation The VW field is dosimetrically superior to the PW beam and can be an efficient method for minimizing acute, late radiation morbidity and reduce the linear accelerator loading by decreasing the radiation delivery time.
Key Words: Breast cancer, Virtual wedge, Physical wedge, Dose distribution
TOOLS
PDF Links  PDF Links
Full text via DOI  Full text via DOI
Download Citation  Download Citation
Share:      
METRICS
1,814
View
35
Download
Related article
Editorial Office
Department of Radiation Oncology, Samsung Medical Center,
Proton Therapy Center, B2, 81, Irwon-ro, Gangnam-gu, Seoul 06351, Republic of Korea
Tel : +82-2-3410-3617
E-mail: rojeditor@gmail.com, roj@kosro.or.kr
Copyright © The Korean Society for Radiation Oncology.                      Developed in M2PI
Close layer
prev next