Journal of the Korean Society for Therapeutic Radiology 1991;9(2): 337-342. |
The Comparison of the KAPM Dosimetric Protocol (1990) with the TG-21 and Clambda/C(E) Method |
Byong Yong Yi, Eun Kyung Choi, Hyesook Chang |
Department of Therapeutic Radiology, Asan Medical Center, College of Medicine, University of Ulsan, Seoul, Korea. |
|
|
|
ABSTRACT |
The Comparison of the KAPM Dosimetric Protocol (1990) with the TG-21 and Clambda/C(E) (ICRU-21 and SCRAD protocol) method is studied. The therapetutic range of radiation (photon 4MV-15MV and electron 6MeV-20MeV)and three kinds of the chambers were used in the water phantom. The Results from TG-21 and KAPM Protocol did not show much differences (less than 1%) throughout the whole energy range; N(D) from KAPM protocol and Ngas from TG-21 showed 0.2% deviation mainly from W/e difference between two protocols. But the results from KAPM protocol (1990) and those from Clambda/C(E) Method showed -1.9+/-0.6%(KAPM protocol is higher) deviation for photom beam and +3.3+/-1% (KAPM protocol is lower) deviation for electron beams. |
Key Words:
Dosimetric Protocol, KAPM Dosimetric Protocol, TG-21 |
|
|
|