Journal of the Korean Society for Therapeutic Radiology 1991;9(1): 123-130.
Design and Dose Distribution of Docking Applicator for an Intraoperative Radiation Thherapy
Sung Sil Chu, Gwi Eon Kim, John J L Loh
Department of Radiation Oncology, Yonsei University College of Medicine, Seoul, Korea.
ABSTRACT
A docking intraoperative electron beam applicator system, which is easily docking in the collimator for a linear accelerator after setting a sterilized transparent cone on the tumor bearing area in the operation room, has been designed to optimize dose distribution and to improve the efficiency of radiation treatment method with linear accelerator. This applicator system consisted of collimator holder with shielded metals and docking cone with transparent acrylic cylinder. A number of technical innovations have been used in the design of this system, this docking cone gives a improving latral dose coverage at therapeutic volume. The position of 90% isodose curve under surface of 8 cm diameter cone was extended 4~7 mm at 12 MeV electron and the isodose measurements beneath the cone wall showed hot spots as great as 106% for acrylic cone. The leakage radiation dose to tissues outside the cone wall was reduced as 3~5% of output dose. A comprehensive set of dosimetric characteristics of the intraoperative radiation therapy applicator system is presented.
Key Words: Intraopertive radiation therapy, Docking applicator, Pentagonal cone, Electron dosimetry
Editorial Office
Department of Radiation Oncology, Samsung Medical Center,
Proton Therapy Center, B2, 81, Irwon-ro, Gangnam-gu, Seoul 06351, Republic of Korea
Tel : +82-2-3410-3617
E-mail: rojeditor@gmail.com, roj@kosro.or.kr
Copyright © The Korean Society for Radiation Oncology.                      Developed in M2PI
Close layer
prev next